| Model 401 NO2 Converter™ |
The Model 401 NO2 Converter (above photo, top), when used in combination with the Model 400 Nitric Oxide Monitor (above photo, bottom), allows measurements of NO, NOx (NOx = NO + NO2) and NO2 (as the difference between NOx and NO) in the range 2-2,000 ppb. The NO2 Converter also contains an internal scrubber for scheduled zeroing of the NO Monitor. The NO2 Converter makes use of a molybdenum ("moly") converter heated to 325 °C to reduce NO2 to NO. Frequencies of switching between NO and NOx measurements and zeroing frequencies are chosen from the NO Monitor™ menu. The instrument package may be set to either always measure NO, always measure NOx, or switch back and forth between measurements of NO and NOx at a specified interval (5 min, 15 min or 1 hour). Choices of zeroing frequency are never, continuous, and at 30 min, 1 hour and 4 hour frequencies. Menu choices of NO/NOx and zero measurement frequencies can be customized to your needs. Specifications
|