Home
About Us
Quote
Products
Comparison
Options
Custom
Calibrations
Trade Ins
Rentals
DewLine
Applications
Literature/Links
GO3 Project
R & D
Research

Feedback
Contact Us

Comparison of 2B Tech Ozone Monitors™

Which ozone monitor should I purchase?

The following table compares the specifications of the various 2B Tech Ozone Monitors and is meant to be helpful for deciding which Ozone Monitor is best for your application.  The table is followed by a discussion of the suitability of the different models to various applications.   If you have questions or need help with selecting the Ozone Monitor best suited to your needs, please do not hesitate to contact us.

Specification
Model
Measurement Principle
UV Absorbance
UV Absorbance
UV Absorbance
UV Absorbance
UV Absorbance
Linear Dynamic Range
0-250 ppm
0-250 ppm
0-100 ppm
0-1000 ppm
0-20 wt%
Resolution
0.1 ppb
0.1 ppb
0.1 ppb
0.01 ppm
0.01 wt%
Precision (1σ for 10-s average; aka rms noise)

1.5 ppb  or
2% of Reading

1.0 ppb or
2% of Reading
2.0 ppb or
2% of Reading
0.01 ppm or
2% or reading
0.01 wt% or 2 % of reading
Accuracy

1.5 ppb  or
2% of Reading

1.0 ppb or
2% of Reading
2.0 ppb or
2% of Reading
0.01 ppm or
2% or reading
0.01 wt% or 2 % of reading
Calibration
NIST Traceable
NIST Traceable
NIST Traceable
NIST Traceable
Iodometric Titration
Limit of Detection (2σ)
3.0 ppb
2.0 ppb
4.0 ppb
0.02 ppm
0.02 wt%
Flow Rate (nominal)
1 L/min
1.8 L/min
1 L/min
1 L/min
1 L/min
Flow Rate Requirement
>0.6 L/min
>1.2 L
>0.6 L/min
>0.6 L/min
>0.6 L/min
Baseline Drift

<2 ppb/day
<5 ppb/year

<1 ppb/day
<3 ppb/year
<3 ppb/day
<6 ppb/year
<0.01 ppm/day
<0.03 ppm/year
<0.02%/day
<0.1%/year
Sensitivity Drift
<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
Measurement Time, Frequency
10 s, 0.1 Hz
2 s, 0.5 Hz
10 s, 0.1 Hz
10 s, 0.1 Hz
10 s, 0.1 Hz
Response Time, 100% of Step Change
20 s
2 points
4 s
2 points
20 s
2 points
20 s
2 points
20 s
2 points
Averaging Times
10 s, 1 min, 5 min, 1 hr
2 s, 10 s, 1 min,
5 min, 1 hr
10 s, 1 min,
5 min, 1 hr
10 s, 1 min,
5 min, 1 hr
10 s, 1 min,
5 min, 1 hr
Ozone Units
ppb, ppm, pphm, µg m-3, mg m-3
ppb, ppm, pphm, µg m-3, mg m-3
ppb, ppm, pphm, µg m-3, mg m-3
ppb, ppm, pphm, µg m-3, mg m-3

ppm, pphm,
µg m-3, mg m-3,
mol %, wt %

Pressure Units
mbar, torr
mbar, torr
mbar, torr
mbar, torr
mbar, torr
Temperature Units
°C, K
°C, K
°C, K
°C, K
°C, K
T and P Corrected
Yes
Yes
Yes
Yes
Yes
Operating Temperature Range
0 to 50 °C;
-20 to 50 °C with low T Mods
0 to 50 °C;
-20 to 50 °C with low T Mods
0 to 50 °C
0 to 50 °C
0 to 50 °C
Operating Altitude
~0-30 km
~0-30 km
~0-30 km
~0-5 km
~0-5 km
Power Requirement;
Supplied by battery or 110/220 VAC Power Pack
12 V dc, 335 mA, 4 watt

12 V dc, 415
mA, 5.0 watt

12 V dc, 290
mA, 3.5 watt
12 V dc, 290
mA, 3.5 watt
12 V dc, 290
mA, 3.5 watt
Size
3.5 x 8.5 x 11 in 9 x 21 x 29 cm
3.5 x 8.5 x 11 in 9 x 21 x 29 cm
4 x 7.8 x 8.5 in
10 x 20 x 22 cm
4 x 7.8 x 8.5 in
10 x 20 x 22 cm
4 x 7.8 x 8.5 in
10 x 20 x 22 cm
Weight
4.7 lb (2.1 kg)
4.7 lb (2.1 kg)
4.3 lb (2.0 kg)
4.3 lb (2.0 kg)
4.3 lb (2.0 kg)
Data Logger Capacity
14,336 lines
14,336 lines
14,336 lines
14,336 lines
14,336 lines
Analog Inputs for Internal Logging of Other Instruments, 0-2.5 V
3 Inputs
3 Inputs
No
No
No
Serial Output, RS232
Yes
Yes
Yes
Yes
Yes
USB Output
External Adapter
External Adapter
Yes
Yes
Yes
Data Transfer Baud Rates
2400, 4800, 19200
2400, 4800, 19200
2400, 4800, 19200
2400, 4800, 19200
2400, 4800, 19200
Analog Voltage Output
0-2.5 V
0-2.5 V
0-2.5 V
0-2.5 V
0-2.5 V
Analog Current Output

Optional
4.-20 mA
Upgrade

Optional
4.-20 mA
Upgrade
4-20 mA
4-20 mA
4-20 mA
Output Ranges
User Defined
Scaling Factor in Menu
User Defined
Scaling Factor in Menu
User Defined
Scaling Factor in Menu
User Defined
Scaling Factor in Menu
User Defined
Scaling Factor in Menu
DewLine™
Yes
Yes
Yes
Yes
Yes
GPS Option
Yes
Yes
No
No
No
Flash Card Option
Yes
Yes
No
No
No
Relay with 2 Setpoints
No
No
Yes
Yes
Yes
Backup Air Pump
Yes
Yes
No
No
No
Flow Meter
Yes
Yes
No
No
No
Upgradable to Dual Beam
Yes
N/A
No
No
No
Federal Equivalent Method (FEM)
Yes
N/A
No
N/A
N/A